DIAMENTOR E2
First Time Operation
12
D651.131.00/05 en
4 First Time Operation
4.1 Setting up the
DIAMENTOR E2
The DIAMENTOR E2 can be set up by means of
the folding stand on the back of the device.
Alternatively, the device can be fixed to the wall or a
plain surface by means of the velcro type ribbon
supplied with the DIAMENTOR E2. To do so, fasten
the self-adhesive back of the ribbon to the back of
the DIAMENTOR E2 and the opposite ribbon to the
wall or surface.
4.2 Connecting the Supply Volt-
age
WARNING
The DIAMENTOR E2 must only be connected
to one of the two ways described below.
Use only the listed original cable from PTW.
Observe the installation notes of the supply
voltage.
WARNING
The system has
no
means of isolation from the
supply voltage. When the mains plug is con-
nected to the power line or the cable is clamped
into the DC supply, the system is powered,
even if the DIAMENTOR E2 is turned off.
The DIAMENTOR E2 can be connected with two
different adapters for supply voltage (refer to fig-
ure 2).
1. Connection with PTW power supply T99052
2. Connection with power limiter T26060 to an
external supply voltage according to IEC 60601-
1/ ANSI/AAMI ES 60601-1 /
CAN/CSA-C22.2.No.60601-1
To connect the DIAMENTOR E2 follow the descrip-
tion of the used adapter.
PTW
Power Supply
T99052
Mains
PTW
DIAMENTOR E2
PTW
Chamber
(110... 230)VAC
1
PTW
DIAMENTOR E2
PTW
Chamber
2
T26060
PTW
PTW T25024-2
PTW T25024-2
24 VDC+/-10%
Double /
Reinforced
insulation
IEC 60601-1
2 MOPP
External
Supply Voltage
Figure 4: DIAMENTOR E2 connecting options