DIAMENTOR E2
Peripherals
22
D651.131.00/05 en
8 Peripherals
8.1 Printer
WARNING
Please read the printer manual before putting
the printer into operation!
The DIAMENTOR E2 may be operated in the
patient environment only if the connected
printer meets the requirements of IEC 60950-1 /
UL 60950-1 and if the printer is operated out-
side the patient environment.
The connected printer meets the requirements
of IEC 60950-1/ UL 60950-1 only if it is oper-
ated with the supplied original power supply.
To obtain a printout of the measured value connect
the printer to the DIAMENTOR
E2 (Figure 3,
item 1).
Switch on the printer first or simultaneously with the
DIAMENTOR E2. Set the print mode to “
Prn on
”
(see section 4.5).
The printout contains the device designation „PTW
DIAMENTOR E2“ and the measured value:
8.2 PC
WARNING
Please read the printer manual before putting
the printer into operation!
Operate the DIAMENTOR E2 in the patient en-
vironment only if the connected PC fulfills the
requirements of the IEC 60950-1 / UL 60950-1
and is installed outside the patient environment.
The PC connected to the device must comply
with the standards IEC
60601-1/
ANSI/AAMI ES 60601-1/
CAN/CSA-C22.2.No.60601-1 if it is to be oper-
ated in the patient environment.
WARNING
Before connecting a PC via the RS232 inter-
face, set the print mode to “
Prn off”
(see sec-
tion 4.5).
Use only the original RS232 interface cable
T25030-3.
It is also possible to connect a PC to the RS232
interface of the DIAMENTOR E2 (socket SERIAL).
When connecting a PC to the DIAMENTOR E2,
please pay close attention to the safety information
in section 2.
A detailed description of the interface is supplied on
request: document number D651.131.1.