
CAL Cl
A1046
Accept with <CAL>
Zero point
Range
Fig. 36: Calibration of zero point
4.
Then press
CAL Cl
A1048
continue with <CAL>
Zero point
Slope
Calibration successful
Fig. 37: Calibration of zero point
5.
Import the result of the calibration into the controller memory
by pressing the key
ð
The controller displays the continuous display again and
operates with the results of the calibration.
Incorrect calibration
Should the result of the calibration lie out‐
side the specified tolerance limits, an
error message appears. In this case the
current calibration will not be applied.
Check the prerequisites for the calibra‐
tion and clear the error. Then repeat the
calibration
Start up
52
Содержание DULCODOS Pool Comfort
Страница 4: ...12 EC Declaration of Conformity 77 13 Index 79 Table of contents 4...
Страница 81: ...81...
Страница 82: ...82...
Страница 83: ...83...