7.3.1.2 Calibrating Amperometric Sensors
Calibrating Amperometric Sensors
The process for calibrating amperometric sensors
is identical with all amperometric measured varia‐
bles.
The process for calibrating amperometric meas‐
ured variables is described throughout based on
the measured variable chlorine [Cl]. All other
measured variables require the same process as
the measured variable chlorine [Cl].
The following measured variables can be cali‐
brated using the process described here:
–
Chlorine
–
Chlorine dioxide
–
Bromine
–
Chlorite
–
Ozone
–
Peracetic acid (PES)
–
H
2
O
2
Calibration combining pH and chlorine
It is imperative that you always calibrate the pH
measurement first before the chlorine measure‐
ment. The chlorine measurement always needs to
be calibrated any further time the pH measurement
is calibrated. Otherwise the chlorine measurement
will be inaccurate.
7.3.1.2.1 Selecting the calibration process for amperometric measured variables
There are two calibration processes available for calibrating the
controller:
n
Calibrating the slope
n
Calibrating the zero point
1.
Continuous display ➨
CAL Cl
A1039
Zero point
Slope
Slope calibration
Calibration of zero point
4.00 mA
Last calibration 31.03. 2013 13:11:11
100 %
Fig. 32: [Chlorine] calibration menu
ð
The calibration menu is displayed.
Selecting the calibration process
Start up
47
Содержание DULCODOS Pool Comfort