CAL pH
Last calibration
A1016
Zero point
Slope
Calibration process
CAL setup
continue with <CAL>
14:26:07
7,00 pH
2 point
56.64mV/pH
06/04/2013
Fig. 24: pH sensor calibration (CAL)
1.
Then press
2.
Rinse the sensor thoroughly with water before drying with a
cloth (pad dry, don't rub)
3.
Immerse the sensor in test container 1 containing buffer solu‐
tion (e.g. pH 7). When doing so, move the sensor gently
4.
Continue with
ð
Calibration is running .
[Please wait!] flashes.
CAL pH
Sensor voltage
Sensor calibration in buffer 1
A1017
Buffer temperature
The stability is:
acceptable
good
continue with <CAL>
25.0 °C
0.1 mV
very good
Fig. 25: Display of the sensor stability achieved
5.
The
[acceptable / good / very good] range is displayed
ð
The black part of the horizontal bar indicates the range
detected.
6.
Continue with
7.
[Buffer detection] e.g. [Manual]: Press and set the buffer
value for buffer 1 to the value of the buffer you are using with
the four arrow keys. Confirm the input of the value with
8.
Remove the sensor from the buffer solution, rinse thoroughly
in water and then dry with a cloth (pad dry, don't rub!)
9.
Continue with
10. Immerse the sensor in test container 2 containing buffer solu‐
tion (e.g. pH 4). When doing so, move the sensor gently
11. Continue with
ð
Calibration is running .
[Please wait!] flashes.
Start up
41
Содержание DULCODOS Pool Comfort