CAL pH
Calibration process Data input
A1024
Zero point
Slope
Last calibration
CAL setup
continue with <CAL>
16:47:32
7,00 pH
59.16 mV/pH
06/05/2013
Fig. 30: pH sensor calibration (CAL)
1.
Continue with
CAL pH
6.88 pH
Slope
at 25.0 °C
A1026
or
-58.07 mV/pH
continue with <CAL>
Asymmetry
at 25.0 °C
Zero point
at 25.0 °C
-6.4 mV
Fig. 31: Selection of the settable parameters
2.
Use the arrow keys to select the required menu item and
press
ð
The entry window appears.
3.
Enter the values for your sensor using the arrow keys and
press
4.
Continue with
Incorrect calibration
An error message appears if the result of the cali‐
bration lies outside the specified tolerance limits. In
this case, the current calibration is not applied.
Check the prerequisites for calibration and clear
the error. Then repeat calibration.
5.
Transfer the result of the calibration into the controller
memory by pressing
ð
The controller displays the continuous display again and
operates with the results of the calibration.