6.3.3 Electrical installation of the sensors
Screw the orange-coloured SN6 plug onto the sensors.
1.
Turn the top section of the sensor a quarter turn anticlock‐
wise and remove it
2.
Loosen the clamping screw on the M12 threaded connector
and feed through the measuring line from the control device
3.
Strip the cable ends, fit the cable ends with cable end
sleeves (⌀ max = 0.5 mm
2
) and connect the cable ends to the
2-wire connection: 1 = plus, 2 = minus
4.
Bring approx. 5 cm of the measuring line into the sensor.
5.
Tighten the clamping screw on the threaded connector
6.
Push the top section of the sensor right into the sensor shaft
and tighten it clockwise up to the stop
6.3.4 Cable Cross-Sections and Cable End Sleeves
Minimum cross-section
Maximum cross-section
Stripped insulation length
Without cable end sleeve 0.25 mm
2
1.5 mm
2
Cable end sleeve without
insulation
0.20 mm
2
1.0 mm
2
8 - 9 mm
Cable end sleeve with
insulation
0.20 mm
2
1.0 mm
2
10 - 11 mm
Electrical installation of the pH sensor
Electrical installation of the chlorine
sensor
A0102
Fig. 17: 2-wire connection
Assembly
29