37
Maxprep™ Liquid Handler | TM509 | Revised 1/19
Maxprep™ Software
3. Below the Description: field are groups of variables whose values can be modified by the
administrator. Touch the
+/–
icon to the left of the group to expand/hide any individual group
of accessible variables.
4. For each variable type, the following are listed:
a.
Name:
A description of the variable that is being set.
b.
Value:
The default value for this variable. If the default value falls outside of the allowable
range for the variable, the value will default to the nearest allowable value within the
range.
c.
Min/Max:
The range of values that are acceptable for this variable. Each method will
have an absolute Min/Max range that is set by Promega. The administrator can narrow
this range if desired to meet laboratory needs. The software will check that the entries
conform to the following conditions:
i. The minimum value is less than or equal to the maximum value.
ii. The minimum and maximum values are within the absolute range set by Promega.
iii. The default value is within the specified range.
If there is an issue with any of these conditions, the software will automatically reset the
offending value to the nearest allowable value within the range.
d.
User Modifiable:
A check box to indicate whether users can change the default value
during a run.
Note:
Not all variables that can be specified in a variant method are available for users
to modify during a method run.
Information regarding the variables that can be set for method types or specific methods,
and the impact of these settings, can be found in Technical Manuals specific to
preprocessing or post-processing methods.
i.
Preprocessing Methods for the Maxprep
™
Liquid Handler Technical Manual #
TM529.
ii.
QuantiFluor
®
Dye Quantitation Methods for the Maxprep
™
Liquid Handler Technical
Manual #
TM531.
iii.
Amplification Setup Methods for the Maxprep
™
Liquid Handler Technical Manual
#
TM526.
iv.
Normalization, Dilution and Transfer Methods for the Maxprep
™
Liquid Handler
Technical Manual #
TM530.
5. The administrator can modify any of the accessible variables to meet the needs of the
laboratory workflow.
Содержание Maxprep AS9100
Страница 2: ......
Страница 6: ...4 Maxprep Liquid Handler TM509 Revised 1 19 Table of Contents...
Страница 16: ...14 Maxprep Liquid Handler TM509 Revised 1 19 Introduction Notes...
Страница 24: ...22 Maxprep Liquid Handler TM509 Revised 1 19 Hardware Overview Notes...
Страница 54: ...52 Maxprep Liquid Handler TM509 Revised 1 19 Maxprep Software Notes...
Страница 74: ...72 Maxprep Liquid Handler TM509 Revised 1 19 Operating the Maxprep Liquid Handler Notes...
Страница 78: ...76 Maxprep Liquid Handler TM509 Revised 1 19 Cleaning the Maxprep Liquid Handler Notes...
Страница 82: ...80 Maxprep Liquid Handler TM509 Revised 1 19 Troubleshooting Notes...
Страница 84: ...82 Maxprep Liquid Handler TM509 Revised 1 19 Instrument Service Notes...
Страница 90: ...88 Maxprep Liquid Handler TM509 Revised 1 19 Certificate of Decontamination Notes...