Page 19
7.3 Calibration on calibration standard
When the instrument indicates that the calibration is outdated, or if it is required by
the applicable regulations, the instrument should be calibrated.
Remove the instrument and the calibration
standard from its carrying case.
Tilt the instrument slightly forward facing and
insert it into the frontal prong support. Tilt the
device back into horizontal position, it should
click into place.
The device will indicate incorrect placement
with an error message.
Press the symbols
and
to open the
calibration function. By default night visibility
(R
L
) is calibrated. Trigger the calibration by
pressing
. The “Calibration Result”
displays the measured value as well as the
deviation from the expected value in percent.
Successful calibration is indicated by
“Calibration OK”. Confirm the calibration by
pressing
and to return to the
measuring mode.