Page 18
7
Calibrate
7.1 Calibration standard
The calibration standard has been measured and the calibration value of R
L
has
been inscribed.
Always protect the calibration standard from dust, moisture and other environmental
influences and safley store it in the carrying case of the Zehntner ZRM 6010.
After the expiry date of the calibration standard a factory calibration of the
ZRM 6014 and its corresponding calibration standard is required. Contact
either Proceqr or your authorized Proceq agent. The factory calibration is valid
for 2 years.
The calibration standard delivered with the Zehntner ZRM 6010 is not
changeable and is valid only for the delivered instrument.
7.2 Cleaning of the calibration standard
If the calibration standard is soiled or covered by dust, it can be cleaned carefully
using window cleaner and a soft tissue.
A damaged or a polluted calibration standard may cause incorrect calibration
and therefore incorrect measuring results.
Calibration standard
Label