Practical example of interval timer
At the interval timer the first measurement starts right after pressing the measuring
button
.
It is recommended to combine the interval timer with the wet timer if the first
measurement should start after a certain period after the measuring button has
been pressed. Please also refer to “Practical sample combination of wet timer and
interval timer”.
Here, the performance of the R
L
-value of a moistened type-II road marking tape is
measured during 60 seconds at an interval of 5 seconds. As the measurements
should start right after moistening the road marking tape, it does not make sense
using the wet timer. This example requires the following settings:
•
Activate the internal timer.
•
Set the interval time to 5 seconds.
•
Set the interval counts to 12.
After carrying out the measurements, the measuring data can be exported to
Microsoft Excel using the software „MappingTools“.For this you need either to store
the measuring data on a USB flash drive as described in chapter 11.6
measuring data on a USB mass storage device
” on page 40 or open the measuring
data directly on the Zehntner ZRM 6010 as described in chapter 12
nd “MappingTools” software” as from page 40.
Example Excel-
report exported using “MappingTools”:
Using this table, you are able to create a diagram to show the R
L
performance.