Page 59
16 Status and error messages
16.1 Status message
– “Battery voltage is critical”
In order to prevent a damage of the battery,
the instrument switches off automatically
before the battery is completely empty. The
Zehntner ZRM 6010 announces this by
displaying the symbol and a status
message.
Charge the battery as described in chapter
14.2
“Charging” on page 57.
16.2 Status message
– “Please calibrate R
L
”
Press the warning triangle to get a more
detailed explanation.
To solve this, carry out the calibration as
described in chapter
” as from page 18. If you are sure
that is not necessary to carry out a
calibration, yo
u can press “Acknowledge all”
and the message will disappear.
To change the calibration interval refer to
chapter 13.6.9
“Calibrate” on page 54.
16.3 Error message
– “Calibration is not OK”
Calibration results with deviations greater
than 15 % are indicated by “Calibration is not
OK”. This can be caused by a soiled or
damaged standard or a soiled optic window.
It is suggested to reject such a calibration
result by pressing
before determining
the reason for the deviation. Clean the
standard and recalibrate again as described
in chapter 0 on page 18 and. By experiencing
repeatedly a deviation greater than 15%, a
calibration by the manufacturer might be
required.