Page 58
15 Maintenance
15.1 Maintenance carried out by the user
You may only carry out the following maintenance and repair yourself:
•
Charging as described in chapter 14.2 on page 57.
•
Cleaning as described in chapter 15.2 on page 58.
•
Replacing the thermal paper roll as described in chapter 10.4
All other maintenance and repair work may only be carried out by Proceq or
your authorized Proceq agent, otherwise all warranty voids.
Make sure that the Zehntner ZRM 6010 is unplugged from power supply and
turned off before any maintenance, except for charging.
The instrument consists of delicate optic and electronic precision parts. Protect
it from shocks, moisture and dust. Please store the instrument in the
transportation case.
15.2 Cleaning
We recommend that the instrument is checked and certified by Proceq or
authorized Proceq Service partner every two years. Otherwise clean the instrument
periodically using water- and oil-free compressed air (max. 1.5 to 2 bar), or wipe the
exit window on the bottom of the optics housing with a soft brush.
The exit window is coated with a special antireflection coating that should not be
damaged. If the soft brush is not sufficient for cleaning the window, you may use a
soft paper tissue and window cleaning liquid.
The optics housing may not be opened under any circumstances since the
measuring geometry would be misaligned. The measuring geometry can only
be adjusted by means of special testing equipment in our factory.