Calibration results with deviations greater
than 15 % are indicated by “Calibration is not
OK”. This can be caused by a soiled or
damaged standard or a soiled optic window.
It is suggested to reject such a calibration
result by pressing
before determining
the reason for the deviation. Clean the
standard as described in chapter 7.2
“Cleaning of the calibration standard” on
page 18 and recalibrate again.
By experiencing repeatedly a deviation greater than 15%, a calibration by the
manufacturer is required.
The data of the calibration settings of R
L
can
be viewed by pressing
,
and
then selecting the geometry. The desired
geometry can be activated/deactivated by
clicking on the line “Calibrate” and choosing
“On” or “Off”.
8
Measure
8.1 Preparation
The Zehntner ZRM 6010 has a telescopic handle which grants triggering
measurements comfortable at the street and in the laboratory.
After removing the Zehntner ZRM 6010 from the carrying case, the length of the
telescopic handle can be adjusted to the right length and the handle can be turned
into the right angle. To change the length, the height shifting knob must be pressed,
to change the angle of the handle and the display, the release turning knob must be
turned. The telescopic handle must click in its final position.
Before the device is put back into the carrying case, the telescopic handle has
to be retracted and the brought into horizontal position.