XTR-PX2-C0006-IN004
4-20
<Fixing the cable>
- Fix the cable under the Rotation Arm with a cable tie (x4).
(To avoid contacting the cover, fix the cables with the cable tie with the lock part at the back side.)
(72)
<Checking operation of the Rotation Arm>
Rotate the Rotation Arm with your hands to check that the cables have no excessive tension.
When the arm has been rotated one turn, check that an extra length of about two fingers is ensured.
short-circuit test
( before AC power cable wiring operation )
AC power cable wiring operation
:
While power switch on the Switch Box is off,
in the presence of an electrician from the facility,
perform the AC power wiring operation.
While turning on the power switch on the Switch Box,
confirm there is no short circuit in the machine by
measuring resistanse between each terminal.
between PE and N: more than 10 kohm
between PE and L: more than 10 kohm
between L and N: more than 10 kohm
Green
White
Blue
Yellow
Cable tie (190 mm)
Cable tie (190 mm)
Cable tie (190 mm)
Cable tie (190 mm)