XTR-PX2-C0006-IN004 leveling w/o Stand
2
Step2
Example of Spacer plate ( in case we use three plates )
Step 3
Fasten tightly the floor screws and wall fixture screws.
Place level measuring tool on the top of Pillar unit.
( shown with red color on the left figure )
Confirm the horizontal levelness is in 0.1 degree.
( X direction , Y direction )
When, we need to adjust the levelness, please use
the Space plate under the Base.
Put the Spacer plate under the Base.