![PreXion PreXion3D Excelsior Скачать руководство пользователя страница 148](http://html1.mh-extra.com/html/prexion/prexion3d-excelsior/prexion3d-excelsior_service-manual_1599689148.webp)
XTR-PX2-C0006-IN004 Ceph option
4-9
10.
Place the red mark so that the circle of it is
overlapped by the Ear rod ring, and then click
[OK].
11.
Check the "Result" field, adjust the
Cephalometric sensor unit tilt according to the
displayed adjustment value.
12.
CEPHALO ARM Tilt Adjustment
Loosen the fixing screws (SCBST8-22) (x3) of CEPHALO ARM BASE, and rotate the adjustment screw
(CP8X30-CM1) of CEPHALO ARM BASE the amount that displayed in the
“Result” field.
Then tighten the fixing screws (SCBST8-22) (x3) again.
Adjustment screw
(CP8X30-CM1)
Fixing screws (SCBST8-22) (x3)