XTR-PX2-C0006-IN004 Ceph option
4-8
7.
Press the e
xposure switch and irradiate X-ray
beam.
8.
After the scanning is complete, click [Vertical
Ear Rod].
→ A screen to check the vertical position of
the Ear Rods appears.
9.
Align with the center of an EAR ROD around
the blue mark.
<When adjustment is not required>
Adjustment is not required when the pin (black
dot) is in the center of the ring (circular ring).
Click [Cancel] and proceed to “4.3
エラー
!
参
照元が見つかりません。
”.
<When adjustment is required>
Adjustment is required when the pin (black dot)
is not in the center of the ring (circular ring).
Proceed to step 10.
* If the positions of the center position of Ear
rod ring and the pin inside EAR ROD are
horizontal, it is OK.
<When adjustment is not required>
<When adjustment is required>
<Example when adjustment is not required >