procedure
step 1:
Remove the nylon clamp and remove cable assembly from the bottom plat
step 2:
Replace the bottom plate to Base STD (Y4150UB-54 ),
by loosening four screws.
step 3:Restore the nylon clamp and cable assembly to the bottom plate.
Caution:
When we don't replace the bottom plate we will have problems such as we cannot put the metal bottom
cover on the Pillar unit, or inside cable can be seen from outside.
Base STD Y4150UB-54 is registered as service parts.
The base place replacement is necessary in the cases below
- when we anchor the new Scanner unit to the floor with using anchor bolt
- when we use the new Scanner unit with the current Stand