XTR-PX2-C0006-IN004 Ceph option
4-16
6.
Click [OK].
7.
The scanner initialization is complete, and wait
for the status indicator changing to green.
→ Exposure instructions are displayed on
the screen.
↓
8.
Press the e
xposure switch and irradiate X-ray
beam.
9.
Perform from step 1 through step 2, and check
that the horizontal position has been adjusted.