XTR-PX2-C0006-IN004
1-7
1.6 Power Requirements (Console)
Connect the Console to a standard commercial power supply.
- PC
: AC100-240V, 50/60Hz, 700W or lower
- Display
: AC100-240V, 50/60Hz, 100W or lower
1.7 Floor load capacity
The floor must be able to withstand the weights of the device (without the Cephalometric Sensor Unit: 165 kg;
with the Cephalometric Sensor Unit: 200 kg), the patient, the operator, and so on.
Recommended load capacity
=
500 kg/m
2