XTR-PX2-C0006-IN004 Ceph option
4-5
4.
Adjustment Work
4.1 Overview of Adjustment Work
From 2.3 through 2.6, it describes how to perform the adjustment work of Cephalometric sensor unit.
Adjustment menus and items are shown below.
Vertical Ear Rod
Adjust the attaching angle of CEPHALO ARM (In most cases, adjustment is
unnecessary).
Check that the ring and the pin are in the same height zones.
Horizontal Radiation field
Adjust the angle of X-ray irradiation beam.
Adjust the positions of the right and right edges of the field.
Vertical Radiation field
Adjust the position (height) of Collimator / aperture.
Adjust the positions of upper and lower edges of the field.
Horizontal Ear Rod
Adjust the angle of attaching position for EAR RODs of Cephalometric
sensor unit.
Check that the ring and the pin are in the same vertical zones.