- 39 -
Chapter
G
Ball Phantom
1
Setting up for ball phantom
Place the ball phantom in the patient positioning mechanism. (Figure 57)
Figure 57
NOTE: Layer must be at 0mm before doing a ball phantom. If layer is not then adjust the thumb wheel till
layer is at 0mm. (Figure 58)
Figure 58
Printed copies of this document are considered uncontrolled.
22811.20.Rev001 07.31.2018
39