- 53 -
2.1.c.i
Too Left/ Too Right
If the shadow ball is too left press the X-Pan collimator pos left
button. (Figure 92)
Figure 92
Press
Test
then
Done
then take another exposure.
If the shadow ball is too right then press the X-Pan collimator
post right button. (Figure 93)
Figure 93
Press
Test
then
Done
then take another exposure.
Printed copies of this document are considered uncontrolled.
22811.20.Rev001 07.31.2018
53