- 77 -
Image should come out as covering top to bottom on the fluorescent screens.
(Figure 141)
Figure 141
3.3.a Too High
Press the y-top limit down and y-bottom limit down. (Figure 142)
Figure 142
Press
Test
and take another exposure. If the beam is aligned then press
Done
and Move onto the Secondary collimator (beaver tail).
Printed copies of this document are considered uncontrolled.
22811.20.Rev001 07.31.2018
77