- 60 -
Place the Ball Phantom into the patient positioning mechanism with the Frankfort
plane alignment tool on top of the Ball Phantom. (Figure 106)
Figure 106
NOTE: Make sure the layer light is on the line of the ball phantom then zero out the layer light.
Press
i
then press
Technical Calibration
(60.0) (Figure 107)
Figure 107
Printed copies of this document are considered uncontrolled.
22811.20.Rev001 07.31.2018
60