- 49 -
2.1.b.ii
Too Right
If the distance from the Center Ball to the right is more than the
left, then move the table toward the column. (Figure 84)
Figure 84
2.1.b.iii
Moving the Table
NOTE: Call Planmeca Technical Support before proceeding.
Loosen the four screws on the table with a 4mm allen. (Figure
85)
Figure 85
NOTE: Screws might be a Torx T25 instead of a 4mm allen
Physically push or pull the table to the left or right.
Printed copies of this document are considered uncontrolled.
22811.20.Rev001 07.31.2018
49