- 88 -
Loosen the left 3mm locking screw and tighten the 3mm right locking screw.
(Figure 185)
Figure 185
Take another exposure. If aligned tighten the left nut and tighten the two
screws on the bottom.
Move onto Beam Check.
5.3 Too Low
Loosen the two 3mm allen screws on the inside of the collimator. (Figure 186)
Figure 186
Printed copies of this document are considered uncontrolled.
22811.20.Rev001 07.31.2018
88