Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline EC X-ray unit E-37
PATIENT POSITIONING LIGHTS
Technical Manual
5.2 Positioning light adjustment
Bending the mirror bracket backwards or forwards will position the light beam.
Figure 68
Rotating the mirror bracket will adjust the light beam in vertical direction.
Figure 69
Adj.beam r
ightleft.eps
Mirror bracket
Adj.beam updo
wn.eps