Chapter B - KEYBOARD FUNCTIONS & MODES
B-16 Planmeca Proline EC X-ray unit
SERVICE MODE KEYBOARD FUNCTIONS
Technical Manual
3.11 Factory preset (recalling factory default quick exposure
settings)
Press the auto key and press and hold the program key for 2 seconds to recall the factory
default quick exposure settings: A 62 kV/5 mA; B 64 kV/6 mA; C 68 kV/7 mA.