Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline EC X-ray unit E-31
PATIENT POSITIONING MECHANISM
Technical Manual
Leave the ball phantom tool in position and place alignment tool adapter to the sensor head.
Place one end of the alignment ruler in the first primary slot (the largest one) and the other
end to the sensor alignment tool.
Figure 56
Loosen the four screws that hold the patient position mechanism in place and adjust its posi-
tion to the left or right until the mechanism is centered.
Figure 57
Make sure that the y-line on the ball phantom is parallel to the line on the underside of the
alignment ruler. Note that the line on the ball phantom and the line on the alignment ruler do
not have to coincide, but they must be parallel, the deviation may be ±1mm.
Figure 58
Tighten the screws, and remove the alignment pin.
Take another ball phantom picture to check the alignment again. Readjust if necessary.
START
END
PLANMECA
No.50971
X-LINE
Y-LINE
Dimax2_adj1.eps
Alignment ruler
Screws
START
END
PLANMECA
No.50971
X-LINE
Y-LINE
line
ca.EPS
Alignment ruler
Lines must not coincide
but they must be parallel
Ball phantom
Содержание Proline EC ceph
Страница 2: ......
Страница 8: ...TOC 6 Planmeca Proline EC Panoramic X ray TABLE OF CONTENTS Technical Manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline EC X ray unit EMC INFORMATION Technical Manual...
Страница 164: ...Chapter E DIMAX3 DIGITAL SYSTEM ADJUSTMENT E 70 Planmeca Proline EC X ray unit IMAGE AREA SHIFTING Technical Manual...
Страница 182: ...Chapter G TROUBLESHOOTING G 12 Planmeca Proline EC X ray unit PROBLEMS IN RADIOGRAPHS Technical Manual...
Страница 188: ...Chapter H PARTS REPLACEMENT REPAIR H 6 Planmeca Proline EC X ray unit CHANGING THE TUBEHEAD Technical Manual...
Страница 205: ...Planmeca Proline EC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 206: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline EC X ray unit Technical Manual...
Страница 208: ......
Страница 209: ......