Chapter D - ADJUSTMENT & CALIBRATION
Planmeca Proline EC X-ray unit D-13
PATIENT POSITIONING MECHANISM
Technical Manual
If the distance from the middle ball to the center of the tenth ball on the left-hand side is less
than 118mm and on the right-hand side more than 122mm then the patient positioning mech-
anism must be moved towards the column.
Figure 16
If the distances from the middle ball to the center of the tenth ball on both the right and left are
more than 122mm then the patient positioning mechanism is too far backward and must be
moved forwards.
Figure 17
If the distances from the middle ball to the center of the tenth ball on both the right and left are
less than 118mm then the patient positioning mechanism is too far forward and must be
moved backwards.
Figure 18
Right-hand side
Film
Less than 118mm
More than 122mm
Film
More than 122mm
More than 122mm
Right-hand side
Less than 118mm
Less than 118mm
Right-hand side
Film
Содержание Proline EC ceph
Страница 2: ......
Страница 8: ...TOC 6 Planmeca Proline EC Panoramic X ray TABLE OF CONTENTS Technical Manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline EC X ray unit EMC INFORMATION Technical Manual...
Страница 164: ...Chapter E DIMAX3 DIGITAL SYSTEM ADJUSTMENT E 70 Planmeca Proline EC X ray unit IMAGE AREA SHIFTING Technical Manual...
Страница 182: ...Chapter G TROUBLESHOOTING G 12 Planmeca Proline EC X ray unit PROBLEMS IN RADIOGRAPHS Technical Manual...
Страница 188: ...Chapter H PARTS REPLACEMENT REPAIR H 6 Planmeca Proline EC X ray unit CHANGING THE TUBEHEAD Technical Manual...
Страница 205: ...Planmeca Proline EC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 206: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline EC X ray unit Technical Manual...
Страница 208: ......
Страница 209: ......