Chapter C - HELP & ERROR MESSAGES
C-2 Planmeca Proline EC X-ray unit
ERROR MESSAGES
Technical Manual
2 ERROR MESSAGES
The purpose of the error messages is to make fault-finding easier. The list of the last 50 error
occurrences is of a great help. Please refer to the section “Error history display” on page B-12
how to display the error list.
ERROR CODE
ERROR MESSAGE EXPLANATION
Er 00
The exposure switch was released too early during exposure.
Er 01
Short loss of power or drop in line voltage.
Er 07
The rotating unit is not in the ready position.
Er 08
The wrong type sensor head (i.e. panoramic type) attached to the cephalostat.
Er 10
Overvoltage in tubehead.
Er 11
Sudden kilovolt drop.
Er 12
Tube filament initializing is not done.
Er 21
Up/down motor time out. The vertical carriage does not reach the limit sensor
within a specified time.
Er 22
Temple rest motor time out.
Er 23
Layer adjust motor time out. The layer mechanism does not reach the limit sensor
within a specified time.
Er 24
Primary slot motor time out. The slot mechanism does not reach the sensor in
specified time.
Er 25
Cassette motor time out. The cassette holder mechanism does not reach the limit
sensor within a specified time.
Er 26
Rotation motor time out. The rotating unit does not reach the limit sensor in the
specified time.
Er 27
Cephalostat scanning mechanism time out.
Er 28
Panoramic sensor head motor time out. The sensor head does not reach the pan-
oramic position in the specified time. Or the safety limit switch is pressed and the
movement of the sensor head is prevented.
Er 30
kV value does not reach given value.
Er 31
mA value does not reach given value.
Er 32
Tube filament control inoperative.
Er 33
Tube filament control inoperative.
Er 40
The rotation does not reach the end limit sensor.
Er 41
Rotation goes over the end limit sensor.
Er 42
Secondary slot not in the position for the panoramic exposure.
Содержание Proline EC ceph
Страница 2: ......
Страница 8: ...TOC 6 Planmeca Proline EC Panoramic X ray TABLE OF CONTENTS Technical Manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline EC X ray unit EMC INFORMATION Technical Manual...
Страница 164: ...Chapter E DIMAX3 DIGITAL SYSTEM ADJUSTMENT E 70 Planmeca Proline EC X ray unit IMAGE AREA SHIFTING Technical Manual...
Страница 182: ...Chapter G TROUBLESHOOTING G 12 Planmeca Proline EC X ray unit PROBLEMS IN RADIOGRAPHS Technical Manual...
Страница 188: ...Chapter H PARTS REPLACEMENT REPAIR H 6 Planmeca Proline EC X ray unit CHANGING THE TUBEHEAD Technical Manual...
Страница 205: ...Planmeca Proline EC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 206: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline EC X ray unit Technical Manual...
Страница 208: ......
Страница 209: ......