Chapter D - ADJUSTMENT & CALIBRATION
Planmeca Proline EC X-ray unit D-33
CEPHALOSTAT ADJUSTMENT
Technical Manual
9 CEPHALOSTAT ADJUSTMENT
9.1 Removing the cassette holder cover
Loosen the two M4x8 BN 5950 bolts on the top of the cover with a 5.5 mm ring spanner.
Loosen the two M4x4 DIN 7985 screws on the bottom of the cover. Pull the cover away from
the cassette holder.
Figure 36
RemCephCo
v
er2.eps
M4x8 BN 5950
M4x4 DIN 7985
Содержание Proline EC ceph