Chapter D - ADJUSTMENT & CALIBRATION
Planmeca Proline EC X-ray unit D-35
CEPHALOSTAT ADJUSTMENT
Technical Manual
9.3 Head support adjustment
Switch the unit off. Remove the cephalostat arm upper cover, attachment casting cover and
attachment mechanism cover.
Make sure that the locking screw of the adjustment flange is loosened (Fig. 38, 1). Adjust the
head support position according to the Fig. 38. Rotate the head support around its vertical
axis to adjust the ear posts in horizontal direction (Fig. 38, 2) and around its horizontal axis to
adjust the ear posts in vertical direction (Fig. 38, 3).
If the cephalostat cassette holder does not run parallel with the primary slot, adjust the cas-
sette holder in vertical direction (Fig. 38, 4).
Figure 38
Fasten the loosened screws and replace the covers.
Remove the head support alignment tools and place the ear posts in the ear post holders.
4
5
6
PLANMECA PM 2002 Proline CC Cephalostat 18x24
No.652232
1
Soft tissue filter adjustment
(see section “Soft tissue filter adjustmen
2
3
4
Содержание Proline EC ceph
Страница 2: ......
Страница 8: ...TOC 6 Planmeca Proline EC Panoramic X ray TABLE OF CONTENTS Technical Manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline EC X ray unit EMC INFORMATION Technical Manual...
Страница 164: ...Chapter E DIMAX3 DIGITAL SYSTEM ADJUSTMENT E 70 Planmeca Proline EC X ray unit IMAGE AREA SHIFTING Technical Manual...
Страница 182: ...Chapter G TROUBLESHOOTING G 12 Planmeca Proline EC X ray unit PROBLEMS IN RADIOGRAPHS Technical Manual...
Страница 188: ...Chapter H PARTS REPLACEMENT REPAIR H 6 Planmeca Proline EC X ray unit CHANGING THE TUBEHEAD Technical Manual...
Страница 205: ...Planmeca Proline EC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 206: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline EC X ray unit Technical Manual...
Страница 208: ......
Страница 209: ......