Chapter C - HELP & ERROR MESSAGES
C-12 Planmeca Proline EC X-ray unit
ERROR MESSAGE EXPLANATIONS
Technical Manual
ER 82 Keyboard processor
The keyboard processor chip watchdog circuit is not activated.
-Change the keyboard processor board
ER 83 Generator processor
The generator processor watchdog circuit is not activated.
- Change the generator processor board
3.10 Errors 90-99 (other errors)
ER 90 Keyboard processor
Keyboard processor EEPROM not activated.
- Change keyboard processor board
ER 91 Keyboard processor
Keyboard processor stack overflow.
- Change the software on the keyboard processor board.
Er 99 Error code generation error
This error indication can only be result of non matching software versions on generator and
keyboard processors or an indication of defective keyboard processor board.
- Check program version numbers and/or replace keyboard processor
Содержание Proline EC ceph
Страница 2: ......
Страница 8: ...TOC 6 Planmeca Proline EC Panoramic X ray TABLE OF CONTENTS Technical Manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline EC X ray unit EMC INFORMATION Technical Manual...
Страница 164: ...Chapter E DIMAX3 DIGITAL SYSTEM ADJUSTMENT E 70 Planmeca Proline EC X ray unit IMAGE AREA SHIFTING Technical Manual...
Страница 182: ...Chapter G TROUBLESHOOTING G 12 Planmeca Proline EC X ray unit PROBLEMS IN RADIOGRAPHS Technical Manual...
Страница 188: ...Chapter H PARTS REPLACEMENT REPAIR H 6 Planmeca Proline EC X ray unit CHANGING THE TUBEHEAD Technical Manual...
Страница 205: ...Planmeca Proline EC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 206: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline EC X ray unit Technical Manual...
Страница 208: ......
Страница 209: ......