Start:
Stop measuring cycle:
run/halt:halt
Choose
mass-MODE:SAMPLE
Choose
cycle-FUNCT:ADJUST
Choose
cycle-MODE:MONO
or
MULTI
Choose
cycle-ADJ-TYP:COARSE
Start
ADJUST
measurement:
run/halt:run
Procedure:
Signal
mon
With successful Adjust the mass
number
MASS
of the measured
channel is updated with the new
value. If unsuccessful it remains
unchanged.
t
Adjust
Peak found
Status message:
After expiration a status message is displayed:
The 3 main parameters
THRESH, RANGE, MASS
can be entered here directly.
You do no have to switch back to the
channels
group.
Symbols:
OK
The adjust was successful
→
→
→
→
Increase
MASS
slightly
←
←
←
←
Decrease
MASS
slightly
↑↑↑↑
Increase Intensity or lower
THRESH
↓↓↓↓
Decrease Intensity
I
Peak too narrow (e.g. parasitic pulse or poor peak shape). Repeat ADJUST.
If unsuccessful: investigate peak shape.
Status code:
Peak width
Intensity
Mass number
MASS
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Bit=1:
Symbol:
too narrow
I
<
THRESH
↑↑↑↑
too high
↓↓↓↓
too low *)
→
→
→
→
too high
←
←
←
←
too low
→
→
→
→
*) and intensity not dropped back to ½
Status code
Symbols