40
BG 805 981 BE (0201) QMG422.oi
IS-TYP
Specify the ion source installed in the QMA.
AXIAL
Axial ion source
CB
Cross-beam ion source
GRID
Grid ion source
SPM
Sputter process monitor ion source
400
SPEC+
Special ion source positive ions
400
SPEC–
Special ion source negative ions
See
TEST-CS 422
Threshold values of the switching functions
LEVEL-A
LEVEL-B TYPE
With
mass-MODE:SAMPLE or
detect-TYPE:PIRANI,PENNING,A-INPUT
1×10
-24
.....
ABS
Threshold value of the switching function A or B
.....9,99×10
+24
HYST
Upper (A) and lower (B) threshold
OFF
Switching function off
If with
TYPE:HYST
: LEVEL-A < 1.1×LEVEL-B this minimum hysteresis is
automatically set.
See AO-MODE: ion counter
→
35
With cycle-FUNCT:ADJUST and cycle-MODE
:
MONO, See below MASS(mass)
In SAMPLE (MID) mode measurement takes place on this mass number during the
measurement time
DWELL
and the average value of the measurement signal is
formed.
The measurement resolution is up to 24 bits (mantissa)
MASS
With
mass-MODE:SAMPLE
0.00 ... max. 2047.99
The maximum value depends on the mass range
Configuration specification of the existing measurement rage (HF generator)
MASS-R
QME-Typ
MASS-R
QMH-Typ
MASS-R
QMH-Typ
100
QME 125-1
128
QMH 400-1
1024
QMH 410-1
200
QME 125-2
512
QMH 400-5
2048
QMH 410-2
300
d.o.+QMA 430
340
QMH 410-3
Operating mode of the electrometer amplifier
MODE
With
detect-TYPE:FARAD
and
SEM
AUTO
Automatic changeover across all measurement ranges, very universal
AUTO-D
Automatic changeover down to the lower search limit
RANGE-L
.
FIX
Manual range selection for fastest measurements
MODE
Controlling interface
CS 422
Console CS 422
ASCII
RS-232-C in ASCII format
BIN
RS-232-C in binary format
MODEM
RS-232-C with modem in binary format
LAN
Arcnet interface
Each interface can switch to itself and thereby interrupt others.
See
TEST-SERVICE
IS-TYP (config)
KEY-T (config)
LEVEL-A, -B (trip)
LOG-DEC (output)
MASS (cycle)
MASS (mass)
MASS-R (config-SYSTEM)
MODE (amplif)
MODE (config-CTRL)
MODE (config)