BG 805 981 BE (0201) QMG422.oi
23
QC
HV+
HV-
A
C
B
QC
HV+ HV–
Cable the system in accordance with the following tables
Short TNC and short SHV = short circuit plugs
Configuration config-SYSTEM-DETECT:FARAD
Module
Connector
Connection
l [m]
Module
Connector
Comments
QC422
QMH
control cable QMH
3
QMH
(QC)
Optional
extension 7m
IS420
QMA
Ion source
3
QMA
IS
or 10m
QMH
RF+
RF–
Radio frequency
0.7
QMA
RF A
RF B
Polarity see
test report
QMH
FA
Field axis
0.7
QMA
FA
EP422
Input
Meas. signal
---
QMA
EP(FARAD)
---
control cable
0.8
QMH
ep1/farad
In addition to 1. Configuration config-SYSTEM-DETECT:SEM
Module
Connector
Connection
l [m]
Module
Connector
Comments
HV42x
HV–
High voltage
3
QMA
HV–
or 10 m
HV421
HV+
short SHV
---
---
---
CD
remains open
---
---
short HV
---
QMA
HV+
EP422
Input
Meas. signal
---
QMA
EP(SEM)
---
control cable
0.8
QMH
ep2/sem
short TNC
---
QMA
EP(FARAD)
if only 1 EP
In addition to 1. Configuration config-SYSTEM-DETECT:CD-SEM
Module
Connector
Connection
l [m]
Module
Connector
Comment
HV421
CD
High voltage CD
3
QMA
CD
or 10 m
HV–
High voltage
3
QMA
HV–
or 10 m
HV +
short SHV
---
---
---
EP422
Input
Meas. signal
---
QMA
EP(SEM)
---
control cable
0.8
QMH
ep2/sem
short TNC
---
QMA
EP(FARAD)
if only 1 EP
4.5 Cabling with QMA 400
1. Faraday cup
2. SEM 217, HV 420 or HV 421
3.
SEM 218 (CD-SEM)