BG 805 981 BE (0201) QMG422.oi
35
The parameters below are listed in alphabetic order by name.
Notation:
Additional information on the utilization, the advantages and disadvantages of
individual settings can be found in Chapter 6.
ADJ-TYP
Type of search with
cycle-FUNCT:ADJUST
COARSE
Coarse search
→
52
FINE
Fine search
→
54.
AI-CH
Only with
detect-TYPE:A-INPUT
0 ... 15
AI 421 channel number to be measured
AO-CH
Not with
detect-TYPE:PIRANI, PENNING, A-INPUT
1 ... 12
Output channel of AO 421 or IC 421 for the measured value of the
selected channel.
NONE
No analog output assigned
In
halt
condition the outputs are set to 0 V, except when they are seized by
computer outputs.
Format selection for analog output to AO / IC 421 and
mon
(
→
55).
Electrometer in Fix-Range:
AO-MODE
amplif-MODE:FIX
and
detect-TYPE:FARAD,SEM,EXTERN
LIN
Linear output in selected measurement
RANGE
LOG 3D
Logarithmic, 3 decades, 3
V / dec. within
RANGE
Electrometer in Auto-Range:
AO-MODE
amplif-MODE:AUTO, AUTO-D
and
detect-TYPE:FARAD,SEM
LIN
Linear output within range selected with
O-RNG
LOG 3D
Logarithmic, 3 decades, 3
V / dec. within range selected with
O-
RNG
LOG 8D*
Logarithmic, 8 decades, 1.25 V / dec. across all ranges
*) With
mass-MODE:SCAN
or
PEAK
and
SPEED <
100 ms/u,
with
STAIR <
10ms/u a changeover to 3 decades occurs automatically
Ion counter:
AO-MODE
LOG-DEC
detect-TYPE:ION-CNT
LIN
-----
Linear output within the range selected with
O-RANGE
LOG
3 DEC
Logarithmic, 3 decades, 3
V / dec. within the range selected
with
O-RANGE.
LOG
10 DEC*
Logarithmic, 10 decades, 1 V / dec. 10
-1
...10
8
cps
*) With
mass-MODE:SCAN
or
PEAK
and
SPEED <
50 ms/u a changeover to 3
decades occurs automatically.
Moving average across measurement cycles
→
51
AVERAGE
Only with
mass-MODE:SAMPLE
and
detect-TYPE:FARAD,SEM,ION-CNT,EXTERN
1
No averaging across measurement cycles
2, 4, 8... ...512, 1024
Number of measurement cycles for forming average
5.3 Parameter list
ADJ-TYP (cycle)
AI-CH (detect)
AO-CH (output)
AO-MODE (output)
AVERAGE (mass)
DETECT (config-SYSTEM
)
Parameter
of the function
possibly subfunction(s)