BG 805 981 BE (0201) QMG422.oi
41
MODE
Measurement cycle mode. The cycle is started/stopped with
run/halt
MONO
Single channel measurement in selected channel
MULTI
Measurement of the channels between
BEGIN
and
END
. Channels that are in
aux-STATE:SKIP
state will be skipped.
MODE
Ion source mode
NORMAL
Normal operation with the parameters defined in the ion source set.
DEGAS
Degas mode. The necessary parameters are entered directly.
Mass scan mode, for details refer to
50
MODE
Not with
detect-TYPE:PIRANI, PENNING
and
A-INPUT
SCAN-N
Normal spectrum from the start mass
FIRST
across the scan width
WIDTH
at the speed set with
SPEED.
SCAN-F
Same, with FIR filter.
STAIR
Spectrum with integer mass jumps
SAMPLE
Measurement on mass
MASS
with averaging across
DWELL
time.
PEAK-L
Peak search (Level criterion) from
FIRST
via
WIDTH
with the speed
SPEED.
Significant data reduction because only the intensities and mass
number of detected peaks are output.
PEAK-F
Same, with FIR filter.
Format of the measured value at the analog output
mon
MONITOR
Not with
detect-TYPE:PIRANI, PENNING
and
A-INPUT
LIN / LOG
See
AO-MODE
RNG-CODE
Range-Code: E-12=1V.. E-5=8V (only for test purposes)
See
TEST-SERVICE
Display next 32 bits with
DIG-IN
and
DIG-OUT
Next error message, if more than one exists.
NODE
Only with
config-CTRL-MODE:LAN
1 ... 255
ARCNET node address
See
TEST-SERVICE
Offset correction, see
FUNCT(cycle)
OPTION
Configuration input for CP 400 ion counter preamplifier
NO
No CP 400
CP
CP 400 exists
Output-Range of AO 421 and
mon
, see
AO- MODE
O-RNG
with
detect-TYPE:ION-CNT
or
amplif-MODE:AUTO, AUTO-D
E-1 ... E+8
In ion counter mode
E-5 ... E-12
In electrometer mode
Measurement pause during channel change (
→
49).
Not with
detect-TYPE:PIRANI, PENNING
and
A-INPUT
.
The actual pause time is displayed if it can be calculated. AUTO is displayed with
amplif-MODE:AUTO
or
AUTO-D
. However,
P-CAL
is still effective.
Press the
PAUSE
soft key to get to the
P-CAL
submenu.
P-CAL
0.0 ... 9.9
Pause time calibration factor
→
49
MODE (cycle)
MODE (ion src)
MODE (mass)
MONITOR (output)
NEG (config)
NEXT (di/do)
NEXT (error)
NODE (config-CTRL)
OFFSET (config)
OFFSET (cycle-FUNCT)
OPTION (config-SYSTEM)
O-RNG (output
)
PAUSE (amplif)
P-CAL (amplif-PAUSE)