44
BG 805 981 BE (0201) QMG422.oi
CS 422
Test of the CS 422 console with:
DISP-T
Endless test of the LC display. After the test has been canceled the Display-
RAM test result is displayed.
EPROM-T
After the EPROM test the result is displayed and the checksum is displayed at
the soft key.
KEY-T
Endless test of the keyboard. Consecutively press all keys to display the
corresponding value.
PRG-NR
The program number of the installed firmware (program version) is displayed.
RAM-T
After the RAM test the result is displayed.
DSP-...
Test of the signal processor: EPROM-T, PRG-NR, RAM-T as above
QMS-...
Test of the system controller:: EPROM-T, PRG-NR, RAM-T as above
SERVICE
Test programs only for factory use
With mass-MODE:PEAK-L and PEAK-F or with cycle-FUNCT:ADJUST
Minimum intensity at which a peak is detected by the peak processor and adjust
algorithm.
THRESH
amplif-MODE:FIX and
detect-TYPE:FARAD, SEM or EXTERN
0.01, 0.03, 0,1... 30 % f.s.d
With Fixrange in % of the full scale deflection
THRESH
amplif-MODE:
AUTO or AUTO-D and
detect-TYPE:FARAD, SEM
1E-15, 1E-14.... 1E- 8
With Autorange in [A]
THRESH
detect-TYPE:ION-CNT
1E0, 1E1.... 1E7
In ion counting mode in counts per seconds [cps]
TRIG
Selection of measurement cycle start
INTERN
Start/Stop is performed via CS 422 or interface.
EXT-AUTO
Start on positive slope of the ext. start signal RUN-IN (see p 11). The cycle
runs until terminated with
halt
or the specified number of measurement
cycles has been attained.
EXT-NORM
Start on positive edge of RUN-IN. The cycle runs as long as RUN-IN is high,
or until it is terminated with
halt
or the number of measurement cycles
specified with
CYCLES
has been attained.
EXT-SNGL
Start on positive edge of RUN-IN. The unit must first be armed with
run
. The
cycle runs until it is terminated with
halt
or the number of measurement
cycles specified with
CYCLES
has been attained.
THRESH (mass)
THRESH (...ADJUST)
TRIG (cycle)