32
BG 805 981 BE (0201) QMG422.oi
After the start (
run
key) of the measurement cycle the measured value of the
momentary measurement channel is displayed. In multichannel mode (
cycle-
MODE:MULTI
) the progression of the channels can be seen (possibly with gaps in
fast processes)
In computer mode the display is updated for monitoring purposes.
Choose the parameter to be entered with one of the 6 softkeys. After the key has
been pressed the parameter value flashes.
If the parameter has only two values (e.g. ON/OFF or ×1/×10) you can change it by
simply pressing the soft key again.
Parameter values are entered or changed via the numeric keypad.
0
...
9
numeric keys
+/–
change sign
.
decimal point
exp
exponential entry
del
cancel and return to
measured value display
Accept new value (Enter)
Change parameters in small increments ...
... and in large increments. When the keys are pressed
continuously the entire value range is scanned. The new value
becomes effective without
The operation is subdivided into four
function groups:
channels
select
detect
mass
amplif
trip
output
aux
di/do
config
error
emiss
V1...V6
V7...
sem hv
ion src
cycle
run/halt
filam
sem
general
ion source
operation
mono
multi
halt
Their function is described in the next Chapter.
Each group comprises several function keys for calling a function (e.g.
mass
in the
channels
group). Each function contains up to 6 parameters.
Measured value display
Softkeys
Number pad
Function groups
cycle mode
start/end channel
actual
channe
l
mass
number
error messages and
warnings
mesured value
numerical ...
... and as
bargraph
associated channel
numbers
switching
functions
scan
mode
other switching
functions
not
activated
OFF
ON