18
BG 805 981 BE (0201) QMG422.oi
16 mm
36 mm
QC
HV+
HV-
86 mm
37
m
m
CP400 Dimension diagram
Slots
2
Number per unit
max. 1
not simultaneously with HV 421
Supply
5 V / 0.4 A ; +24 V / 0.15 A ; -24 V / 0.05 A
Fuse F1
→
65
SEM voltage HV-
0 ... -3500 V
Resolution 1 V, ripple <10 mV
pp
Admissible load
15 M
Ω
Current limitation <1 mA
Internal resistance
620 k
Ω
Settling time
0.8 s to 0.1%
HV connector
SHV
Signal ground from QMA via HV
cable
HV test connection
1 V pro 1 kV
R
i
= 2 k
Ω
Potential isolation
0.5 V
max
Between chassis and QMA-GND
Weight
0.42 kg
Slots
3
Number per unit
max. 1
not simultaneously with HV 420
Supply
5 V / 0.2 A ; +24 V / 0.2 A ; -24 V / 0.2 A
HV connector
SHV
SEM voltage HV–/HV+
0 (-750
1
)...-3500 V
Resolution 1 V, ripple <10 mV
pp
Admissible load
17 M
Ω
Current limitation 0.8 mA
Settling time
<0.7 s
to 1%
CD voltage to GND
-6300 V
Adjustable -4.7 ... -6.4 kV
Ripple <10 mV
pp
Admissible load
100 M
Ω
Current limitation < 0.5 mA
Bias voltage HV–
+3.1 / -3.1 kV
For
SPEC– / SPEC+
adjustable 2.4...3.2 kV
Test terminals
1 V per 1 kV
CD Test also for HV–
Weight
1.7 kg
1)
In
SPEC+
mode
3.9 HV 420
High voltage supply
3.10 HV 421 High voltage
supply