Start-up
Differential circuit breaker.
In case of insulation defect, for personnel protection you must install on the
main power supply a type B differential circuit breaker GFI (or RCD) of 30 mA
minimum.
This equipment protection device is compatible with type T.T electrical
network.
For other network type T.N or I.T, apply the right protection device. Contact
adixen product manufacturer for advice.
In all cases, comply with current local regulations.