Once all of the methanol has evaporated, it’s time to reassemble the system. Insert the
crystal holder into the bore of the bearing being careful to line up the groove with the
ball in the crystal holder.
11.
Gently press down on the crystal holder until it stops. You may need to twist the holder
a small amount from side to side to seat it on the retracted plunger.
Do not force it
,
you may damage the stop contact.
12.
Using another swab, clean any debris from the knurled stop collar. Notice the black
spots in the image below.
13.
Screw the knurled collar on all the way then back off one and a half (1.5) full
revolutions. This will set the lower limit of travel for the crystal.