16 . Spotlight 150 User's Guide
Location and Ventilation
The Spotlight 150 is installed by a PerkinElmer Service Representative, who will be able to
advise on the positioning of the system. To allow for adequate cooling, the system should
not be sited near to room heating equipment, for example, central-heating radiators. There
should be a minimum gap of at least 15 cm (6 inches) from the top and side surfaces of the
microscope to permit adequate cooling.
WARNING
Make sure that the switch at the electrical supply inlet at the rear of the
microscope is not obstructed.