Techniques for Collecting Spectra . 61
Collecting a Spectrum from a Thick Sample
If your sample is very thick, you will have to lower the stage to focus the sample. For very
thick samples you may have to remove the lower cassegrain assembly; then only the
reflectance method can be used to view the sample and collect spectra.
To remove the lower cassegrain and lower the stage
1.
Remove any sample from the stage.
2.
Click
Zero
in the Correction section of the Setup Microscope tab.
The lower cassegrain moves to the position where the infrared beam is focused if no
sample is on the stage.
3.
Move the sample stage to its highest possible position.
4.
Release the locking lever located on the right side on the back of the lower cassegrain
assembly (Figure 22).
Figure 22 The lower cassegrain
5.
Gently slide the cassegrain assembly forward and out of the dovetail connector.
NOTE:
To make withdrawal easier, pull the locking lever gently.
6.
Click
Park
in the Correction section of the Setup Microscope tab.
The lower cassegrain bracket moves down. Park is the lowest position to which the
lower cassegrain bracket can move when the lower cassegrain is fitted.
The button text now updates to
Lower Park
.
7.
Click
Lower Park
again.
The lower cassegrain bracket moves to its lowest position.
8.
Lower the stage.
Содержание Spotlight 150
Страница 1: ...Spotlight 150 User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 11: ...Warnings and Safety Information ...
Страница 23: ...Overview of the Spotlight 150 ...
Страница 32: ...32 Spotlight 150 User s Guide ...
Страница 33: ...Getting Ready to Use the Spotlight 150 ...
Страница 45: ...Preparing Samples ...
Страница 58: ...58 Spotlight 150 User s Guide ...
Страница 59: ...Techniques for Collecting Spectra ...
Страница 94: ...Maintenance ...
Страница 102: ...Appendices ...