88 . RamanFlex 400 Series Getting Started Guide
Servicing the Spectrometer
We recommend that you do not remove the protective housing or external panels of the
spectrometer module except to clean the air filters (See
Cleaning the Air Filters
on page 87).
There are no other user-serviceable parts in the instrument. If you require an instrument
service please contact PerkinElmer. Only PerkinElmer Service Engineers should attempt to
service the spectrometer.
WARNING
Removal of the protective housing or exterior panels on any instrument
will allow potentially hazardous Class 3B laser radiation to be emitted
and will invalidate the instrument warranty.
Maintenance and Service of the Fiber Optic Probe
The FOP is designed to be maintenance free and should not be opened or adjusted by
anyone other than a PerkinElmer Service Engineer. Contact PerkinElmer if you have any
questions regarding maintenance or servicing of the FOP.
The immersion sleeve (if purchased) is manufactured with 316 stainless steel or Hastelloy C
and contains either a quartz or sapphire optical window that is sealed. The quartz window
can be removed for cleaning by unscrewing the end of the immersion tube. The sapphire
window cannot be removed as it is welded into place. If removing the quartz window, do not
lose the PTFE O-rings, as these are required to maintain the immersion seal. Without these
PTFE rings the immersion tube will leak and the FOP will be damaged.
The front lens or window on the fiber probe body and the immersion sleeve window should
be left free from debris. To clean simply wipe the fiber faces clean with alcohol soaked cotton
swabs or lint-free tissues.
Damage to the FOP as a result of accident, neglect, misuse, or as a result of service or
modification by anyone other than PerkinElmer will invalidate the FOP warranty.
NOTE: For details of how to maintain and clean a triggered fiber optic probe, see the
Raman
Triggered Fiber Optic Probe
leaflet (L1321887).
Содержание RamanFlex 400
Страница 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 27: ...Raman History Theory and Instrumentation ...
Страница 34: ...34 RamanFlex 400 Series Getting Started Guide ...
Страница 35: ...Preparing your Facility ...
Страница 40: ...40 RamanFlex 400 Series Getting Started Guide ...
Страница 41: ...Installing your RamanFlex 400 ...
Страница 61: ...Introduction to the RamanFlex 400 ...
Страница 68: ...68 RamanFlex 400 Series Getting Started Guide ...
Страница 69: ...Using the RamanFlex 400 ...
Страница 85: ...Maintenance and Troubleshooting ...