Raman History, Theory and Instrumentation . 29
Raman Theory
When photons from a monochromatic light source impinge on a sample, a small fraction are
scattered in all directions. Some of the scattered photons are of the same energy as those of
the incident beam, and hence the same frequency, giving rise to Rayleigh scattering
Figure 5 The Raman effect: Irradiation of sample and subsequent scatter
Some photons will be inelastically scattered and will be detected at either longer wavelengths
(lower frequencies,
ν
’, Stokes scattering) or shorter wavelengths (higher frequencies,
ν
’’,
Anti-Stokes
scattering). With Stokes scattering, some of the incident photons give up a
portion of energy to the molecules and are detected at a lower frequency (longer
wavelength) than the Rayleigh radiation (Figure 6).
Figure 6 Rayleigh, Stokes and Anti-Stokes Scattering
Wavenumber/ cm
-
1
O
No change in energy
no useful information
Shift from excitation line
-
in
wavenumber
/ cm
-
1
Int
e
ns
it
y
Wavenumber/ cm
-
1
O
Shift from excitation line
-
in wavenumber/cm
-1
Loss of energy
spectra (Stokes)
Photons gain energy
but, signal is very small,
(anti-Stokes)
Laser Photons in
Sample
Lens
Laser Photons in
Lens
Virtual Energy State
Rayleigh
Anti-Stokes
Stokes
Excited Energy
ν
= 0
ν
= 0
ν
= 0
Содержание RamanFlex 400
Страница 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 27: ...Raman History Theory and Instrumentation ...
Страница 34: ...34 RamanFlex 400 Series Getting Started Guide ...
Страница 35: ...Preparing your Facility ...
Страница 40: ...40 RamanFlex 400 Series Getting Started Guide ...
Страница 41: ...Installing your RamanFlex 400 ...
Страница 61: ...Introduction to the RamanFlex 400 ...
Страница 68: ...68 RamanFlex 400 Series Getting Started Guide ...
Страница 69: ...Using the RamanFlex 400 ...
Страница 85: ...Maintenance and Troubleshooting ...